Brief Introduction of Structure
The internal test room is made by high quality SUS304 stainless steel; heat insulating material: high density
fiberglass wool; insulation material thickness: 80mm. The heat inside the internal room may not be conducted
to the outside, to keep balanced and stable temperature inside the internal room;
lOne transparent window is located at a proper position on test chamber door, to obverse the change of
specimen inside the internal room. The window is made by multi-layered tempered glass. The internal binding conductive coating is transparent, heat insulation, steam and
frost resistant and other advantages.
The mixing system adopts long axle fan motor, multi-blades wheel, made of stainless steel with good temperature resistance. The air is circulated forcedly vertically
in convection, to keep even and stable temperature inside the internal room;
The door and chamber body are sealed by high elastic double layer seal, with high temperature resistance.
The door inner plate is made by stainless steel. The door adopts non-reaction door handle, easy operation;
The test hole (at the left side of the machine) can be used for connecting power cable or signal cable (hole
diameter and number can be decided by the user).
The cooling system is installed at the bottom. The compressor is "Tecumseh", imported from France. The
single cooling system can guarantee that the temperature or humidity reaches the setting value within the
stipulated time.
The bottom of the test chamber is equipped with high quality PU moving wheels, which can be fixed. The
test chamber can be moved conveniently to the place required , finally fixed.
The control system is at the right side of the test chamber, including temperature controller and control
switches etc. convenient and easy operation and repair.
The air sampling system includes: sampling tube, two 100ml suction bottles, gas sampling pump, gas
flowmeter.
Implementation Standard:
ASTMD6007-02: standard test method for determination of formaldehyde concentration in wood products by small scale environment box ENV717-1: formaldehyde emission measurement environment box method of artificial board EN13419-1: the method for determining the release of VOCs for building products in the first part of the test environment box |
Reference standard: ASTMD5116-97: standard guide for the determination of organic release of indoor materials and products in small scale environment boxes ASTMD6330-98: "standard operation for the determination of VOCs (deformaldehyde-formaldehyde) release in wooden panel under test conditions" ASTMD6670-01: standard operation for measuring the release of indoor materials and VOCs in a full-scale environment box ASTME1333-96: standard test method for determination of formaldehyde concentration and release rate in wood products by large environment box |
Inner volume
|
(1.0±0.02)m³
|
Test temperature range
|
(10-40)ºC, accuracy:±0.5ºC, Resolution 0.1ºC, Adjustable
|
test humidity adjustable range
|
(30-80)%, accuracy:±3%, Resolution 0.1%, Adjustable
|
Air exchange
|
(0.1-2)m/s, accuracy:±0.05 m/s, Adjustable
|
Air speed
|
(0.2-2)m³/h, accuracy: ±5%, Adjustable
|
Leakproofness
|
1KPa relative pressure, leakage rate<10-3×1m³/min, import and export air differ<1%
|
background levels
|
formaldehyde≤0.006mg/m³
|
Adiabaticity
|
The wall and door of the climate box is effectively thermal insulation
|
Noise
|
less than 60dB
|
Continuous working hours
|
not less than 40 days
|
monitoring parameter
|
Monitoring the temperature and humidity, air discharge, switching time, working time, data can be transferred and stored
|
Method of controlling humidity
|
The dew point control method is adopted to control the relative humidity, the humidity is stable, and the fluctuation range
<3%.rh. no water appear on the inside |
Out dimension (W×D×H)
|
1630 x 1200 x 1850mm
|